STANDARD OPERATING PROCEDURES for OPERATION OF AUTOCLAVE - ROOM 320
STANDARD OPERATING PROCEDURES
- Using control box on wall to the right of the autoclave, rotate dial to "on" position.
Allow 15 minute warm up.
- Open autoclave door by rotating wheel to the left.
- Load autoclave.
- Close door and lock into place by rotating wheel to the right.
- Using the numbers on the keypad on the control box to the right of the autoclave, choose
the appropriate cycle:
- Cycle #1 - Wrapped pack (30 min. sterilize/30 min. dry)
- Cycle #2 - Wrapped pack (30 min. sterilize/20 min. dry)
- Cycle #3 - Flash/unwrapped cycle (3 min sterilize/0 min. dry)
- Cycle #4 - Liquids (30 min. sterilize/0 min. dry)
Press appropriate cycle number then "enter". If you push the wrong cycle number for
the desired cycle, do not press "enter"; press "clear". To abort cycle, press the
"clear" key at any point in the cycle.
- A special cycle may be run by pressing "load" and setting the parameters desired. Once
the cycle is completed, it will not be stored in memory.
- At end of day, turn autoclave knob on key box to "off".